Function:
1. Numerical calculation: calculate systolic blood pressure, mean pressure, diastolic blood pressure, and pulse rate.
2. Automatic measurement mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90 minutes.
3. Manual measurement mode: the user needed to manually start a measurement.
4. Calibration mode: the value of continuity in the provision of cuff pressure
5. Measurement unit: mmHg
6. Serial Communication: RS232 port, TTL-level, 4800 baud rate
7. Self checking: self-checking function available after reset
Characteristics:
Ø Small in size, simple, excellent reliability and ease of installation clinical monitoring of the accuracy; Excellent reliability and accuracy of clinical monitoring
Ø The latest digital signal processing technology to achieve anti-noise and software algorithms anti-interference by moving arm;
Ø Capable of accomplishing all NIBP measurement items;
Ø Comply with CE safety design;
Ø Comply with all NIBP standard of ANSI/AAMI;
Ø Applies to adults, children and Newborn;
Ø 5V/12V single powersupply to avoid interference with the digital circuit by the pump
Ø Fully compatible with the protocol of CAS Non-invasive blood pressure module (NIBP)
Specifications:
Measuring range
|
Adult:
30 to 255 mmHg systolic blood pressure
An average of 20 to 235 mmHg pressure
15 to 220 mmHg diastolic blood pressure
Pulse rate of 30 to 240 times / min
Newborn:
30 to 135 mmHg systolic blood pressure
An average of 20 to 125 mmHg pressure
15 to 110 mmHg diastolic blood pressure
Pulse Rate of 40 to 240 times / min
|
Measurement time
|
Adults: up to more than 120 seconds;
Newborn: up to more than 90 seconds;
|
Restriction on software over-voltage
|
Adult: Less than 290mmHg;
Newborns: no more than 145mmHg;
|
Protection on hardware over-voltage
|
Adult: 300 ±10 mmHg;
Newborn: 150 ±4 mmHg;
|
Measuring principle
|
Oscillometry
|
Deflation approach
|
ladder Deflated
|
The initial inflation pressure value
|
Adult: 150 mmHg;
Newborn: 70 mmHg;
|
Sensor zero reset
|
Prior to each measurement
|
Resolution
|
1 mmHg
|
Measurement accuracy
|
Maximum average deviation: no more than ± 5 mmHg;
maximum standard deviation: less than 8 mmHg;
|
Pressure accuracy instructions
|
± 3 mmHg
|
Leakage
|
less than 6 mmHg Per minute;
|
Electrical performance
|
Voltage: 12 VDC (adapt to the corresponding valve), 5 VDC
Quiescent current: 80 mA
Measurement current: 200 mA
inflated current: 600mA
|
Environment
|
Operating Temperature: 0 to 55 Celsius degrees
Storage Temperature: -20 to 65 Celsius degrees
Humidity Range: 0 to 95%, non-condenser
Operation Altitude: -500 m to 4600 meters
Life Span: can operate 100,000 blood pressure measurements
|
Size
|
12.6 cm x 6.8 cm x 2.8 cm
|
Weight:
|
180 g
|
|