I. FEATURES
n Complies with IEC 60601, higher security designation with dual CPUs;
n CE certified;
n Two optional modes, drip-rate mode and volumetric mode;
n High brightness LCD display and user-friendly operation interface;
n Easy switch among d/min, mL/h and volume/time units;
n Available for Bolus mode;
n Anti-reverse function on motor to prevent upstream;
n Self-check when start up and in process;
n Top pressure sensor for preventing underflow;
n Pressure and air-in-line detection devices both with 8 sensitivities setting, dynamically display pressure value;
n IV set management function: Up to 20 IV sets can be added, stored and deleted in memory, convenient for operation after calibration;
n Adjustable 8 volume levels of alarm for convenient operation in different environments;
n Multi-Languages: English, French, Turkish, Polish, Spanish, Russian, Chinese.
II. SPECIFICATIONS
Specification
|
Infusion Mode:
|
Drip-rate and Volumetric
|
Pumping Mechanism:
|
Peristaltic finger pump
|
IV Sets:
|
Standard IV sets of 10, 15, 20 and 60d/mL
|
Optional Rate Unit:
|
d/min, mL/h and volume/time
|
Flow Rate Range:
|
10, 15, 20 d/mL IV sets: 1.0~1200 mL/h
|
60 d/mL IV set: 1.0~150.0 mL/h
|
Increment:
|
0.1mL/ h
|
Drip Rate Range:
|
10 d/mL IV set: 1.0~200 d/min
|
15 d/mL IV set: 1.0~300 d/min
|
20 d/mL IV set: 1.0~400 d/min
|
60 d/mL IV set: 1.0~150 d/min
|
Volume Limit:
|
1~9999 mL
|
Volume Infused Display:
|
0~9999 mL & >9999 mL
|
Flow Rate Accuracy:
|
≤ ± 5% (after calibration)
|
Drip Rate Accuracy:
|
≤ ± 1.5%
|
Purge Rate:
|
600 mL/h
|
KVO Rate:
|
1 mL/h
|
Bolus Mode:
|
Bolus rate adjustable
|
Occlusion Pressure:
|
Pressure detection device with 8 sensitivities setting,
40 ~ 140KPa.
|
Air-in-line Detection:
|
Air-in-line detection device with 8 sensitivities setting,
50 ~ 1000μL.
|
IV Set Management
&
Flow Rate Calibration:
|
Up to 20 IV sets of customers' brand can be added, stored and deleted in memory,
convenient for operation after calibration
|
Display:
|
Blue LCD with high brightness
|
Power Source:
|
AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA
|
Battery:
|
Rechargeable Ni-MH battery: DC12V,2000mAh,
life: ≥ 4 hours at 25mL/h
|
Classification:
|
ClassⅡ, CF, IPX4
|
Dimensions and N.W.:
|
150mm x 150mm x 235mm; 2.3kg
|
Warranty:
|
Two years
|
Accessory
|
Drip Sensor
|
1 pc
|
Power Cord
|
1 pc
|
Operator Manual:
|
1 pc
|
Packing
|
Inner box:
|
230*175*285mm
|
Carton:
|
480*360*300 (4 sets/CTN)
|
|