C500 Mainstream ETCO2 Module

C500 Mainstream ETCO2 Module

 The Mainstream CO2 Sensor C500 is designed specifically for mainstream measurements using sophisticated infrared absorption spectroscopy to make EtCO2 monitoring simple and accurate, allowing for flexibility to monitor all patient (Adult/paediatric and infant ) in all environments.

 

With mainstream, measurements are taken directly at the patients' airway, so response is faster and there is less chance of erroneous, artifact data. The analyzer portion of the sensor cannot be contaminated with patient secretions, and there are no disposable components to C500. And the maintenance requirements and overall cost-of-ownership are minimized. Our exclusive family of airway adapters are easy to use and durable enough even for long-term intubated patients.

 

The Mainstream CO2 Sensor C500 can be integrated with other components, software packages and accessories, making it easy for you to configure your patient monitoring systems with state-of-the-art respiratory management and analysis technology.

 

Key features

 

• compatible with Respironics CapnoStat5 or Phasein’s IRMA

 

• Warm-up time: 3 seconds full specification; Direct measurement without time delay

 

• Advanced technology of condensation compensation

 

• Long-life infrared detector

 

• No disposable accessory and airway adaptor can be used up to 100 times.

 

overall cost-of-ownership are minimized

 

 

Model C500 Mainstream CO2 SENSOR - SPECIFICATIONS

 

 

Transducer Type

Mainstream

 

 

Principle of Operation

Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.

Initialization Time

Capnogram displayed in less than 3 seconds,

 

at an ambient temperature of 25, full specifications within 30 seconds

CO2 Measurement Range

0 to 152 mmHg, 0 to 20.0%, 0 to 20 kPa

CO2 Calculation Method

BTPS (Body Temperature Pressure Saturated)

CO2 Rise Time (10 - 90% of step change of final CO2 value)

 

8 ms - for detector

 

50 ms - for system

CO2 Resolution

0.01 mmHg 0 to 69 mmHg, 0.02 mmHg 70 to 150 mmHg

CO2 Accuracy *

0 - 40 mmHg

±2 mmHg

 

41 - 70 mmHg

±5% of reading

 

71 - 100 mmHg

±8% of reading

 

101 - 150 mmHg

±10% of reading

 

* NOTE: Temperature at 35 .

CO2 Stability

Short Term Drift: Drift over four hours shall not exceed 2 mmHg maximum.

 

Long Term Drift: Accuracy specification will be maintained over a 120 hour period.

CO2 Noise

RMS noise of the sensor shall be less than or equal to 0.25 mmHg at 7.5% CO2

Sampling Rate

100 Hz

 

 

Respiration Rate Range

2 to 150 breaths per minute (BPM)

Respiration Rate Accuracy

±1 breath

 

 

Calibration

No routine user calibration required. An airway adapter zero is required when insert an airway adapter.

Safety lock-outs:

• System does not allow adapter zero for 20 seconds after the last breath is detected.

 

• System does not allow adapter zero if temperature is not stable.

 

• System does not allow adapter zero if there is CO2 gas in the airway adapter.

ETCO2 Calculation

Method: Peak of the expired CO2 waveform

 

Selections: 1 breath, 10 second, 20 second

 

Note: the minimum reported differential value between the baseline and the CO2 value shall be 5 mmHg.

Inspired CO2 Measurement

Range: 3 to 50 mmHg

 

Method: Lowest reading of the CO2 waveform in the previous 20 seconds

Compensations (Host Controlled)

 

 

 

Compensations for: Expired O2 , balance gas (N2 , N2 O, He) and anesthetic agents

 

Uses gas compensation information and barometric pressure to correct the raw carbon dioxide value

O2 Compensation

Range: 0 to 100%

 

Resolution: 1%

 

 

 

Default: 16%

 

 

N2O Compensation

Range: 0 (off) or 1 (on)

 

Default: Off

 

 

 

Note: If ON, the balance of the mixture is O2

He Compensation

Range: 0 (off) or 1 (on)

 

Default: Off

 

 

 

Note: If ON, the balance of the mixture is O2

Total Pressure

Range: 400-850 mmHg.

 

Total pressure = Barometric plus Airway pressure. Barometric pressure provided by host.

Airway Adapter Type

Pediatric/Adult Reusable Airway Adapter, Infant/Pediatric Reusable Airway Adapter

Input Voltage

5.00 Volts (± 5%)

 

Power Consumption

900mW (typical)

 

Warm-up Time

3 seconds

 

 

Interface

Highly configurable serial digital interface(TTL/RS232), CO2 waveform data, Breath Detect Algorithm data

Temperature and Humidity

Operating: 0 to 55ºC, 10 to 90% RH, non-condensing

 

Storage: -40 to 70ºC, <90% RH, non-condensing

Dimensions

50mm*22mm*36mm

Weight

< 95 grams, only the module

Interconnection

Standard Lemo Redel 8-pin plastic Material PSU

 

Pin out:1. VA 5.0V; 2. Shield Shield; 3. DGND Digital return; 4. VA 5.0V; 5. TxD Serial data from the Module;

 

6. RxD Serial data from Host; 7. AGND Analog return; 8. NC