Transducer Type
|
Mainstream
|
|
|
Principle of Operation
|
Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.
|
Initialization Time
|
Capnogram displayed in less than 3 seconds,
|
|
at an ambient temperature of 25℃, full specifications within 30 seconds
|
CO2 Measurement Range
|
0 to 152 mmHg, 0 to 20.0%, 0 to 20 kPa
|
CO2 Calculation Method
|
BTPS (Body Temperature Pressure Saturated)
|
CO2 Rise Time (10 - 90% of step change of final CO2 value)
|
|
8 ms - for detector
|
|
50 ms - for system
|
CO2 Resolution
|
0.01 mmHg 0 to 69 mmHg, 0.02 mmHg 70 to 150 mmHg
|
CO2 Accuracy *
|
0 - 40 mmHg
|
±2 mmHg
|
|
41 - 70 mmHg
|
±5% of reading
|
|
71 - 100 mmHg
|
±8% of reading
|
|
101 - 150 mmHg
|
±10% of reading
|
|
* NOTE: Temperature at 35 ℃.
|
CO2 Stability
|
Short Term Drift: Drift over four hours shall not exceed 2 mmHg maximum.
|
|
Long Term Drift: Accuracy specification will be maintained over a 120 hour period.
|
CO2 Noise
|
RMS noise of the sensor shall be less than or equal to 0.25 mmHg at 7.5% CO2
|
Sampling Rate
|
100 Hz
|
|
|
Respiration Rate Range
|
2 to 150 breaths per minute (BPM)
|
Respiration Rate Accuracy
|
±1 breath
|
|
|
Calibration
|
No routine user calibration required. An airway adapter zero is required when insert an airway adapter.
|
Safety lock-outs:
|
• System does not allow adapter zero for 20 seconds after the last breath is detected.
|
|
• System does not allow adapter zero if temperature is not stable.
|
|
• System does not allow adapter zero if there is CO2 gas in the airway adapter.
|
ETCO2 Calculation
|
Method: Peak of the expired CO2 waveform
|
|
Selections: 1 breath, 10 second, 20 second
|
|
Note: the minimum reported differential value between the baseline and the CO2 value shall be 5 mmHg.
|
Inspired CO2 Measurement
|
Range: 3 to 50 mmHg
|
|
Method: Lowest reading of the CO2 waveform in the previous 20 seconds
|
Compensations (Host Controlled)
|
|
|
|
Compensations for: Expired O2 , balance gas (N2 , N2 O, He) and anesthetic agents
|
|
Uses gas compensation information and barometric pressure to correct the raw carbon dioxide value
|
O2 Compensation
|
Range: 0 to 100%
|
|
Resolution: 1%
|
|
|
|
Default: 16%
|
|
|
N2O Compensation
|
Range: 0 (off) or 1 (on)
|
|
Default: Off
|
|
|
|
Note: If ON, the balance of the mixture is O2
|
He Compensation
|
Range: 0 (off) or 1 (on)
|
|
Default: Off
|
|
|
|
Note: If ON, the balance of the mixture is O2
|
Total Pressure
|
Range: 400-850 mmHg.
|
|
Total pressure = Barometric plus Airway pressure. Barometric pressure provided by host.
|
Airway Adapter Type
|
Pediatric/Adult Reusable Airway Adapter, Infant/Pediatric Reusable Airway Adapter
|
Input Voltage
|
5.00 Volts (± 5%)
|
|
Power Consumption
|
900mW (typical)
|
|
Warm-up Time
|
3 seconds
|
|
|
Interface
|
Highly configurable serial digital interface(TTL/RS232), CO2 waveform data, Breath Detect Algorithm data
|
Temperature and Humidity
|
Operating: 0 to 55ºC, 10 to 90% RH, non-condensing
|
|
Storage: -40 to 70ºC, <90% RH, non-condensing
|
Dimensions
|
50mm*22mm*36mm
|
Weight
|
< 95 grams, only the module
|
Interconnection
|
Standard Lemo Redel 8-pin plastic Material PSU
|
|
Pin out:1. VA 5.0V; 2. Shield Shield; 3. DGND Digital return; 4. VA 5.0V; 5. TxD Serial data from the Module;
|
|
6. RxD Serial data from Host; 7. AGND Analog return; 8. NC
|