Portable Automated External Defibrillator (AED7000)

Portable Automated External Defibrillator (AED7000)

Standard Configuration

1 Main Unit

1 Pair of Pads

2  Batteries

1  Handbag

 

Features

1.Three-step defibrillation process

2. Two-button operation

3. Extensive voice and visual prompts for the operator

4. Biphasic energy output

5. Lock-out protection to prevent inadvertent defibrillation

6.Continuous event recording for reporting each use to a printer or computer

7.Weekly self-test to ensure readiness

 Specifications:

Output

Biphasic Truncated Exponential

Energy Sequence

150, 150, 200J

Charge Time

8 sec. to 150J; 10 sec. to 200J (New Battery)

Analysis Time

9 sec.

The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery

 Less than 30 seconds

The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks

Less than 35 seconds

The maximum time from initially switching power on to readiness for discharge

Less than 25 seconds

Audible Prompts

22 audible prompts

Visual Prompts

LED prompts

Controls

Two buttons-ON/Off, Shock

Output Energy Accuracy

±15% into any impedance from 25 to 175Ω

The Maximum Voltage

1200±50V

Output disabled when the patient impedance is outside limits

20Ωto 200Ω